When treating large tumors, a conventional solid radiation beam can damage surrounding healthy tissue. Doctors came up with the idea of splitting the beam into many thin streams, like water from a watering can with fine holes. This approach is called spatially fractionated radiation therapy, or GRID therapy.
To apply GRID therapy, you need to calculate precisely how the beams will travel through the patient's body. For this, special computer programs, treatment planning systems, are used. The authors were the first to configure such a system, called RayStation, for GRID therapy. They used a lead plate with holes, a GRID block, placed in front of the patient, which turns a solid beam into a grid of thin rays.
Accuracy was checked using special detectors that measure dose at various points. The researchers also developed a way to standardize the setup of the GRID block in RayStation, so other clinics can replicate the method without lengthy trial and error. This is a step toward making treatment of large tumors safer and more predictable.
as written by the authors arXiv:2604.03967
Spatially Fractionated Radiation Therapy (SFRT) or GRID therapy has been in existence for more than a century. GRID therapy was invented as an approach to address the challenges posed by bulky tumors. To treat patients with GRID therapy, the protocol needs to be implemented within the Treatment Planning System (TPS) software. Also, the dose needed for treatment must be validated. A few TPS software vendors such as Elekta and Varian have protocols for treating patients with GRID therapy. However, we are the first to implement the protocols within RayStation TPS. To achieve these, we used the .decimal GRID block applicator to create conformal treatment plans. For the beam modeling, several dosimetry data, and measurements such as percentage depth dose (PDD), beam profile, output, and GRID factors, were obtained for different field sizes and energies. The dose distribution from the GRID openings covered the Planning Target Volume (PTV), with the QA plan having a 98% agreement with the planned dose. Also, we used the Mapcheck2 results to verify the adjustments made to the script variables such as aperture size (hole and spacing). We developed a robust way to validate the dose needed for patient treatment using the GRID block applicator. Also, we have proposed a novel approach to standardize the clinical implementation of the GRID block in RayStation TPS.
GRID therapy has been around for over a century, but only now is it being integrated into modern computer-based treatment planning systems.