Noble liquids such as liquid xenon emit in the vacuum ultraviolet (127 nm for argon, 175 nm for xenon) and are widely used in searches for dark matter and neutrinoless double beta decay. Silicon photomultipliers (SiPMs) are replacing photomultiplier tubes due to their lower radioactivity and better resolution, but their performance depends strongly on temperature. Commercial characterization systems are not designed for cryogenic temperatures and controlled VUV illumination, and the large areas of detectors require automated testing of hundreds of devices. This work stands apart: neighboring projects in the archive concern pocket-sized coolers for quantum circuits, transparent electrodes, and lamps for atomic deposition, but none address the task of mass testing SiPMs in the vacuum ultraviolet at cryogenic temperatures.
The cryostat is built around a liquid nitrogen scheme with two cooling modes: high-power (direct contact with a cold finger) and low-power (heat transfer through attenuator rings). Aluminum rings and supports form a low-pass filter that smooths temperature fluctuations, while polyimide-based heaters allow the temperature to be raised up to 340 K. VUV light is delivered through a 142 mm diameter CaF2 window built into a sealed dark box flushed with dry nitrogen to remove moisture. A two-axis motorized system based on stepper motors moves the light source over the sample with an accuracy of 0.1 mm. Light sources include a xenon scintillation lamp with radioactive 252Cf to create single photons and a pulsed xenon lamp with bandpass filters for a calibrated beam. The electronics include a multiplexer for measuring currents down to 100 fA and a fast channel with an amplifier up to 2.4 GHz.
Temperature stability reached 1.2 mK in a sliding 10-minute window after the fill transient and 13.8 mK over a full fill cycle. Allan deviation drops to 0.14 mK with 2-minute averaging. For ten SiPMs from the Hamamatsu array, we measured dark count (0.6 counts/s/mm² at 3 V overvoltage), single photoelectron charge (0.378 pC), excess charge from correlated avalanches (0.050 PE), and photon detection efficiency (16.3%). The temperature coefficient of breakdown voltage was (52.8±0.5) mV/K. The VUV light beam had a radius of 1.09×1.21 mm and fit entirely on one SiPM, allowing PDE measurement with accuracy limited by photodiode calibration.
This work differs from neighboring projects in that it combines cryogenic stabilization with external VUV optics and automated scanning, which is absent in pocket coolers for quantum circuits or lamps for atomic deposition. It demonstrates for the first time mass testing of SiPMs under conditions approaching a liquid xenon detector, with precise beam positioning and flux calibration. This fills the gap between single laboratory measurements and industrial quality control.
The setup can be extended to test large SiPM arrays with an area of about 100 cm², to study radiation damage from VUV exposure, and to build a model of correlated avalanches with temperature dependence. Integration with other light sources and automation of the full characterization cycle are possible.
The results will directly influence the development of detectors for nEXO, DarkSide-20k, and other liquid noble gas experiments, which require thousands of SiPMs with known characteristics.
Plans include measuring the radiation hardness of SiPMs under prolonged VUV exposure and characterizing large-area arrays at various temperatures.
In detector physics, an open question remains about the long-term stability of SiPMs in liquid xenon and the contribution of correlated noise to energy resolution. This setup allows systematic study of these effects under controlled conditions.
This section is addressed to the author of the paper and their colleagues. Neighbouring works were found by meaning-based search across our archive, not by word overlap. We do not review: we show the field around the work.
What the work does
The paper describes a cryostat with a CaF2 window for automated testing of SiPMs in the vacuum ultraviolet at temperatures of 120-340 K. Current-voltage characteristics and the photodetection efficiency of a Hamamatsu array were measured at 165 K, and a temperature stability of 1.2 mK and a positioning accuracy of 0.1 mm were demonstrated.
Where its strength is
The setup combines cryogenic stabilization, external VUV optics, and automated scanning, which makes it possible to test SiPMs in large batches under conditions close to those of liquid xenon experiments.
Where it could go next
Worth trying
Why this may matter
Such a platform could become a standard tool for selecting and certifying SiPMs in experiments with liquid noble gases, where large-scale testing at cryogenic temperatures and under VUV illumination is required.
Nearby works in our archive
The archive is sparse around this work: the closest review is only 0.55, against a usual 0.64. It looks like few have drilled nearby.
The calcium fluoride window transmits 88% of light at 172 nm, but it is so hygroscopic that it must be constantly flushed with dry nitrogen, otherwise moisture from the air absorbs the ultraviolet light.